2 login/password protection, Login/password protection – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1249
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.2 (ProcessLab)
■■■■■■■■
1233
■
Default reasons
Definition of reasons for signing and modifying methods, determinations
and sample data.
6.2.2.2
Login/Password protection
Tab: Configuration
▶ Tools ▶ Security settings ▶ Login/Password protection
On the tab Login/Password protection is defined whether the user must
log in with his/her name or with name and password, and how the password
is to be monitored and constructed.
Settings according to 21 CFR 11
[Set]
Sets the parameters to default settings for FDA-compliant work.
Login policies
Enforce login with user name
If this check box is enabled then every time that the program starts, the
login window will appear in which the user must enter his/her user name.
If this check box is disabled no login is necessary and the user logged in
under Windows will be adopted as user name. In this case all subsequent
parameters are deactivated.
Selection
on | off
Default value
off
FDA default: on
Enforce login with password
If this check box is enabled then every time that the program starts, the
login window will appear in which the user has to enter a password as well
as his/her user name. If this option is disabled, then all subsequent param-
eters are deactivated.
Selection
on | off
Default value
off
FDA default: on
Password monitoring by tiamo
If this option is enabled then the password will be monitored by tiamo
according to the following parameters.
Selection
on | off
Default value
off
FDA default: on