Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 476
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
460
■■■■■■■■
tiamo 2.2 (ProcessLab)
related to the total measured value alteration, even small changes in the
measured value can be evaluated as a break-point for a small measured value
range.
Range
0 ... 1.0
Default value
0.3
Slope
Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.
Range
0.0 ... 10.0
Default value
0.9
Smoothing factor
The higher the smoothing factor, the fewer break-points will be found.
Range
2 ... 20
Default value
5
Window
A range (window) can be defined on the measured value axis, on the volume
axis or on the time axis. The break-point evaluation will only be carried out
in the defined window. Only the first break-point in the defined window will
be recognized.
Selection
Measured value | Volume | Time | off
Default value
off
Lower limit
Value for the lower limit of the window.
Window = Measured value
Range
-20.000 ... 20.000 pH
Default value
-20.000 pH
Window = Volume
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Window = Time
Range
0 ... 999999 s
Default value
0 s
Upper limit
Value for the upper limit of the window.