5 met ipol - stop conditions, Stop conditions – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 627
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
611
tab under Sensor to automatic or continuous, then the temperature will
be measured continuously. This value is used for temperature correction in
pH measurements.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
25.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
25.0 °C
5.6.3.3.4.5
MET Ipol - Stop conditions
Tab: Method
▶ MET Ipol ▶ Properties... ▶ Stop conditions
Command name
Name of the command.
Input
25 characters
Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop the
titration.
Stop volume
Stops when the given volume has been added after the start of the titration
(including start conditions). The stop volume should be adapted to suit the
sample weight or the titration vessel size.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
100,000 mL
Selection
off
Titrino
Range
0.000 ... 9999.99 mL
Default value
100.00 mL
Selection
off
Stop measured value
Stops when the preset value for a measuring point has been exceeded or
not achieved since the start of the titration.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Selection
off
Default value
off