6 tet - thermometric evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 886
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
870
■■■■■■■■
tiamo 2.2 (ProcessLab)
Range
0 ... 999999 s
Selection
off
Default value
off
Stop slope
Stops when the criterion defined for the first derivative of the measured value
(slope) is fulfilled.
First field: Selection of the reference condition.
Selection
< | >
Default value
<
Second field: Definition of reference value for the derivative of the measured
value.
Range
-999999.999 ... 999999.999 °C/mL (Increment:
0.001)
Default value
0.100 °C/mL
Selection
off
Additional volume after stop
After stopping the titration, this additional volume is dosed according to the
stop conditions defined. The curve shape after reaching the endpoint can
also be seen this way.
Range
0.01 ... 50.00 mL
Default value
0.5 mL
Selection
off
Filling rate
Rate (volume/time unit) at which the buret is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the dosing unit
used. If the selected filling rate is higher than the maximum possible dosing
rate, it will be automatically corrected to this value.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.9.6
TET - Thermometric evaluation
Tab: Methode
▶ TET ▶ Properties... ▶ Thermometric evaluation
Command name
Name of the command.
Input
25 characters
Parameters for the thermometric evaluation of the titration curve with auto-
matic endpoint recognition.