6 tet - thermometric evaluation – Metrohm tiamo 2.2 (ProcessLab) User Manual

Page 886

background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

870

■■■■■■■■

tiamo 2.2 (ProcessLab)

Range

0 ... 999999 s

Selection

off

Default value

off

Stop slope

Stops when the criterion defined for the first derivative of the measured value
(slope) is fulfilled.

First field: Selection of the reference condition.

Selection

< | >

Default value

<

Second field: Definition of reference value for the derivative of the measured
value.

Range

-999999.999 ... 999999.999 °C/mL (Increment:
0.001)

Default value

0.100 °C/mL

Selection

off

Additional volume after stop

After stopping the titration, this additional volume is dosed according to the
stop conditions defined. The curve shape after reaching the endpoint can
also be seen this way.

Range

0.01 ... 50.00 mL

Default value

0.5 mL

Selection

off

Filling rate

Rate (volume/time unit) at which the buret is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the dosing unit
used. If the selected filling rate is higher than the maximum possible dosing
rate, it will be automatically corrected to this value.

Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.9.6

TET - Thermometric evaluation

Tab: Methode

TET Properties... Thermometric evaluation

Command name

Name of the command.

Input

25 characters

Parameters for the thermometric evaluation of the titration curve with auto-
matic endpoint recognition.