837
[Properties]
Opens the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 880).
[Delete]
Deletes the external measured value selected in the table.
5.6.3.8.6.10
STAT pH - Fix endpoint evaluation
Dialog window: Method
▶ STAT pH ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation
▶ [New]/[Properties] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-20.000 ... 20.000 pH
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.8.7
STAT U
5.6.3.8.7.1
STAT U - Overview
Dialog window: Method
▶ STAT U ▶ Properties... ▶ STAT U - 'Command name'
Command for STAT titrations with measured value U.
Devices
This command can be executed with the following devices:
Titrando: 835, 836, 842, 857, 902, 906, 907
Titrino: 718, 736, 751, 799
Robotic Titrosampler: 855
Appearance
The command has the following appearance: