5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1004
■■■■■■■■
tiamo 2.2 (ProcessLab)
Solution
Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.
Input
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
Note
If an existing solution is selected as the solution to be added then the
Concentration and Unit of this solution will be adopted for calculating
the standard addition. If not defined is selected, then Concentration
and Unit must be entered on the Standard addition tab.
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando, 867
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1
Sensor
Selection of a sensor of type ISE electrode from the sensors configured in
the sensor table.
Selection
'Sensor name' | ISE electrode
Default value
ISE electrode
Temperature measurement
Type of temperature measurement.