5 dos ph - monitoring, Monitoring – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1098
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1082
■■■■■■■■
tiamo 2.2 (ProcessLab)
Stop time
Stops when the preset time has elapsed since the start of the dosing.
Range
0 ... 999999 s
Selection
off
Default value
off
off
Means that no stop will take place.
Filling rate
Rate at which the dosing cylinder is to be filled after dosing. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing
device, it will automatically be reduced during dosing to the largest possible
value.
Titrando, 855
Range
0.01 ... 166.00 mL/min
Default value
maximum mL/min
Titrino
Range
0.01 ... 150.00 mL/min
Default value
maximum mL/min
Note
The filling rate should be reduced for viscous liquids.
5.6.6.3.4.5
DOS pH - Monitoring
Tab: Method
▶ DOS pH ▶ Properties... ▶ Monitoring
Command name
Name of the command.
Input
25 characters
The following monitoring processes can be switched on and defined on this
tab:
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this option is switched on then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.