5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
570
■■■■■■■■
tiamo 2.2 (ProcessLab)
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.3.2.10
MET pH - Volume window
Dialog window: Method
▶ MET pH ▶ Properties... ▶ Potentiometric evalua-
tion
▶ Evaluation with volume window (mL) ▶ [New]/[Properties] ▶ Volume
window #
Volume windows are defined sub-regions (windows) on the volume axis for
which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-
tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.
Lower limit
Lower limit for the volume window.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
0.00000 mL
Upper limit
Upper limit for the volume window.
Titrando, 855
Range
0.00000 ... 9999.99 mL
Default value
9999.99 mL
EP criterion pH
Criterion for the recognition of endpoints. Endpoints whose discovered ERC
value is smaller than the value entered here will not be recognized.
Titrando, 855
Range
0.1 ... 9.99
Default value
0.50
EP recognition
Filter for the recognition of equivalence points:
Titrando, 855
Selection
first | greatest | last | ascending | descending
Default value
first