5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
856
■■■■■■■■
tiamo 2.2 (ProcessLab)
Titrino
Range
0.00 ... 150.00 mL/min
Default value
0.00 mL/min
Lower hysteresis
The lower hysteresis represents a tolerance range for the lower limit of the
dosing rate. When the dosing rate exceeds the lower limit by this hysteresis
value, the action Dosing rate lower limit ok is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.20 mL/min
Upper limit
Upper limit of the dosing rate. If the dosing rate exceeds this limit, then the
event Dosing rate upper limit exceeded will be triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
166.00 mL/min
Titrino
Range
0.00 ... 150.00 mL/min
Default value
150.00 mL/min
Upper hysteresis
Upper hysteresis of the dosing rate. When the dosing rate again falls below
the upper limit by this hysteresis value, then the event Dosing rate upper
limit ok is triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.20 mL/min
Action
Selection of the event to be triggered if the upper or lower limit is exceeded:
Selection
Stop determination | Stop command (Titrando,
855) | Wait for [Continue] | Wait for limit ok |
none
Default value
none
Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.