5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
762
■■■■■■■■
tiamo 2.2 (ProcessLab)
Selection
Sensor name | Metal electrode
Default value
Metal electrode
U(pol)
The polarization voltage is the voltage applied to the polarized electrode
during an amperometric measurement.
Titrando
Range
-1250 ... 1250 mV (Increment: 25)
Default value
400 mV
Titrino
Range
-1270 ... 1270 mV (Increment: 10)
Default value
400 mV
Electrode test
on | off (Default value: off)
If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.
Titrando
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.