5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
871
Evaluation start
Possibility to postpone the start of the evaluation. The evaluation starts not
before having added the solution volume.
Range
0.0 ... 50.0 mL
Default value
0.1 mL
Endpoints
Selection of the sorting for the display of the endpoints in the table.
Selection
Sort by volume (ascending) | Sort by peak
height (descending)
Default value
Sort by volume (ascending)
Table of endpoints
A maximum of 9 endpoints can be defined. The parameters for determining
these endpoints are displayed in the window table which cannot be edited
and which contains the columns Reaction type and EP criterion.
[New]
Opens the dialog window Endpoint # in which the parameters for a new
endpoint can be entered (see Chapter 5.6.3.9.9, page 874).
[Properties]
Opens the dialog window Endpoint # in which the parameters for the
selected endpoint can be edited (see Chapter 5.6.3.9.9, page 874).
[Delete]
Deletes the endpoint selected in the table.
5.6.3.9.7
TET - Additional evaluations
Tab: Method
▶ TET ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this option is activated, then the associated values for the other measured
quantities will be interpolated with a fixed measured quantity (Measured
value or Volume) for the fixed endpoint from the measuring point list.