4 kft ipol - control parameters, Control parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 761
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
745
Pause 2
Pause 2
Waiting time, e.g. for the electrode to settle down after the start or a reac-
tion time after the addition of the start volume. The pause follows at the
end of all the start conditions.
Range
0 ... 999999 s
Default value
0 s
5.6.3.5.2.4
KFT Ipol - Control parameters
Tab: Method
▶ KFT Ipol ▶ Properties... ▶ Control parameters
Command name
Name of the command.
Input
25 characters
Endpoint
Ep with
Measured value for the endpoint.
Range
-2000.0 ... 2000.0 mV
Default value
250.0 mV
Selection
off
Titration rate
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical, fast titrations.