5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
741
not defined
No tests will be carried out.
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando (without 852)
Selection
1 | 2
Default value
1
852
Selection
2
Default value
2
Sensor
Selection of a sensor of type Metal electrode from the sensors configured
in the sensor table. The calibration data for the sensor will be adopted for
the determination.
Selection
Sensor name | Metal electrode
Default value
Metal electrode
I(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.
Titrando, 855
Range
-125.0 ... 125.0 µA (Increment: 0.5)
Default value
5.0 µA
Titrino
Range
-127 ... 127 µA (Increment: 1)
Default value
5 µA
Electrode test
on | off (Default value: off)
If this option is enabled, then an electrode test will be carried out for polar-
izable electrodes during the transition from an inactive condition to a
measurement. A check is also made while doing so to ensure that the elec-
trode is properly connected and that no short-circuit is present.
Temperature measurement
Type of temperature measurement.