5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

990

■■■■■■■■

tiamo 2.2 (ProcessLab)

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded

Range

200.0 ... 1100.0 nm

Default value

400.0 nm

Integration time

Integration time for absorption measurement.

Range

0.01 ... 600000 ms

Default value

6 ms

Averaged spectra

The number of spectra that are recorded and averaged.

Range

1 ... 10000

Default value

10

Smoothing

Number of neighboring pixels with which the light intensity is determined
for each measurement pixel.

Range

1 ... 100 Pixels

Default value

1 Pixels

Measurement

Measuring time

Duration of the absorption measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Range

0.1 ... 999999.0 s

Default value

2.0 s

Stop measured value

Stops when the preset measured value has been reached since the start of
the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off