4 stdadd man - measuring parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1015
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
999
Addition solution
Concentration
Concentration of addition solution.
Range
0.001 ... 999999.999
Default value
1.000
Device
Concentration unit of addition solution.
Selection
mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L
Default value
ppm
5.6.4.15.4
STDADD man - Measuring parameters
Tab: Method
▶ STDADD man ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement with drift control
The measurement is stopped as soon as the Signal drift or the Stop
measured value is reached or a defined Waiting time has elapsed.
Measured value drift
The measured value is only accepted if the signal drift defined here has been
fallen short of.
Range
0.1 ... 999.0 mV/min
Default value
10.0 mV/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time