7 kft upol - conditioning, Conditioning – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 787
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
771
Stop time
Titrando
Range
0 ... 999999 s
Selection
off
Default value
off
off
Stops when the preset time (including start conditions) has elapsed since
the start of the titration.
Filling rate
Rate at which the dosing cylinder is to be filled after titration. The maximum
filling rate depends on the cylinder volume of the exchange unit or dosing
device used.
Titrando
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Titrino
Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
5.6.3.5.3.7
KFT Upol - Conditioning
Tab: Method
▶ KFT Upol ▶ Properties... ▶ Conditioning
Command name
Name of the command.
Input
25 characters
Conditioning
Conditioning
on | off (Default value: off)
Switching conditioning on/off. If conditioning is deactivated all following
parameters are not visible.
Start drift
The determination can only be started when the currently measured drift is
less than the start drift entered here. In such cases, the Status in the
Run subwindow displayed on the Single determination or Determina-
tion series tab will then switch to COND READY.