5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
1079
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos
and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for Titrinos.
5.6.6.3.4.3
DOS pH - Dosing parameters
Tab: Method
▶ DOS pH ▶ Properties... ▶ Dosing parameters
Command name
Name of the command.
Input
25 characters
The volume to be added, the dosing rate and the dosing time are defined
on this tab.
Dosing criterion
Selects the two parameters that can be preset for dosing. The third para-
meter is calculated automatically from the formula Volume = dosing time
* dosing rate and is not shown (see Chapter 5.6.6.3.2, page 1068).
Selection
Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time
Default value
Volume/dosing rate
Volume
Fixed volume that is to be dosed. This parameter is only displayed for Dosing
criterion = Volume/Dosing rate or Volume/Dosing time.
Range
0.00000 ... 99999.9 mL
Default value
10.0000 mL