7 det u - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 500
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
484
■■■■■■■■
tiamo 2.2 (ProcessLab)
5.6.3.2.3.7
DET U - Additional evaluations
Tab: Method
▶ DET U ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional methods for evaluation of titration curves can be
activated and defined on this tab:
Fix endpoint evaluation
Fix endpoint evaluation
on | off (Default value: off)
If this check box is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fixed measured quantity (Meas-
ured value, Volume or Time) for the fix endpoint from the measuring
point list.
Moves the selected line up (changes the sequence).
Moves the selected line down (changes the sequence).
[New]
Opens the dialog window Fix endpoint evaluation # to enter a new fixed
endpoint (see Chapter 5.6.3.2.3.11, page 491).
[Properties]
Opens the dialog window Fix endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.3.2.3.11, page 491).
[Delete]
Deletes the selected line.
pK/HNP evaluation
pK/HNP evaluation
on | off (Default value: off)
If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.10.1, page 875).