11 met ph - fix endpoint evaluation, 3 met u, 1 met u - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 587: Met u
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
571
first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be recog-
nized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.
5.6.3.3.2.11
MET pH - Fix endpoint evaluation
Dialog window: Method
▶ MET pH ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation
▶ [New]/[Properties] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Volume
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-20.000 ... 20.000 pH
Time
Range
0.0 ... 999999.9 s
Volume
Range
0.00000 ... 9999.99 mL
5.6.3.3.3
MET U
5.6.3.3.3.1
MET U - Overview
Dialog window: Dialog window
▶ MET U ▶ Properties... ▶ MET U - 'Command name'
Command for Monotonic Equivalence Point Titrations with potentio-
metric voltage measurement.