4 met ipol - titration parameters, Titration parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 624
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
608
■■■■■■■■
tiamo 2.2 (ProcessLab)
Range
0 ... 999999 s
Default value
0 s
5.6.3.3.4.4
MET Ipol - Titration parameters
Tab: Method
▶ MET Ipol ▶ Properties... ▶ Titration parameters
Command name
Name of the command.
Input
25 characters
Parameters defining the run of the titration.
Titration rate
The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.
Selection
slow | optimal | fast | User
Default value
optimal
slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwanted
EPs.
optimal
Parameter set for all standard titrations; optimized for the most frequent
applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration rate.
Measured value acceptance
Note
Will be displayed only if the User option is selected in the Titration
rate list box.
Measured value drift
Drift for the measured value acceptance during the titration.