5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1110
■■■■■■■■
tiamo 2.2 (ProcessLab)
778, 789, 814, 815, 855, 864, 874
Selection
1 | 2 | 3
Default value
1
730, 774
Selection
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12
Default value
1
Titrino
Selection
internal D0
736, 751, 758, 799
Selection
internal D0 | external D1 | external D2
Default value
internal D0
Solution
Input of a solution name or selection of a solution from the solutions defined
in the solution table. If intelligent exchange or dosing units are used, then
a check will be made in the method sequence to see whether the correct
solution has been set on the selected dosing device and whether the dosing
device type is correct. With non-intelligent exchange or dosing units, only
the cylinder volume is checked. At the start of the command, a check is made
of the working life, the validity of the titer and the GLP test interval for the
selected solution.
Input
24 characters
Selection
'Solution name' | not defined
Default value
not defined
not defined
No tests will be carried out.
5.6.6.6
EMPTY
5.6.6.6.1
EMPTY - Overview
Dialog window: Method
▶ EMPTY ▶ Properties... ▶ EMPTY - 'Command name'
Command for Emptying cylinder and tubing of a dosing unit.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 842, 851, 852, 857, 859, 888, 890,
901, 902, 904, 905, 906, 907
Titrino: 736, 751, 758, 799
Dosing Interface: 846
Conductometer: 856