5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
998
■■■■■■■■
tiamo 2.2 (ProcessLab)
5.6.4.15.3
STDADD man - Standard addition
Tab: Method
▶ STDADD man ▶ Properties... ▶ Standard addition
Command name
Name of the command.
Input
25 characters
The parameters for the standard addition can be defined on this tab
Number of additions
Number of times the standard solution is to be added.
Range
1 ... 19
Default value
3
Volume auxiliary solution
Volume of the auxiliary solution (e.g. buffer) which was added to the sample
before the first standard addition.
Range
0 ... 9999.999 mL
Default value
0 mL
Addition mode
Selection
Fixed volume increments | Variable volume
increments
Default value
Fixed volume increments
Fixed volume increments
If this option is enabled a fixed volume increment must be added for each
standard addition.
Variable volume increments
If this option is enabled the defined volume increment must be added
for each standard addition.
Addition volume
Size of the fixed addition volume (only when Addition mode = fixed vol-
ume increments).
Range
0.00000 ... 99999.9 mL
Default value
100 mL
Addition volume 1…19
Size of the individual addition volumes (only when Addition mode = var-
iable volume increments).
Range
0.00000 ... 99999.9 mL
Default value
100 mL