4 stdadd auto - measuring parameters – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1031
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
1015
Delta U
Potential difference that is to be achieved by a standard addition.
Range
1 ... 999 mV
Default value
10 mV
5.6.4.17.4
STDADD auto - Measuring parameters
Tab: Method
▶ STDADD ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
The parameters for the measuring procedure can be defined on this tab.
Measurement with drift control
The measurement is stopped as soon as the defined Signal drift or the
Stop measured value is reached or a defined Waiting time has elapsed.
Measured value drift
The measured value is only accepted if the signal drift defined here has been
fallen short of.
Range
0.1 ... 999.0 mV/min
Default value
10.0 mV/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:
Range
0 ... 999999 s
Default value
52 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.