Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 870
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
854
■■■■■■■■
tiamo 2.2 (ProcessLab)
Monitoring Measured value
Monitoring Measured value
on | off (Default value: off)
If this check box is activated, then the measured value will be monitored and
any exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is triggered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
-2000.0 mV
Titrino
Range
-2000 ... 2000 mV
Default value
-2000 mV
Lower hysteresis
Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value lower
limit ok is triggered.
Titrando, 855
Range
0.0 ... 2000.0 mV
Default value
0.2 mV
Upper limit
Upper limit of the measured value. If the measured value exceeds this limit,
then the event Measured value upper limit exceeded is triggered.
Titrando, 855
Range
-2000.0 ... 2000.0 mV
Default value
2000.0 mV
Titrino
Range
-2000 ... 2000
Default value
2000
Upper hysteresis
The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.