Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 768
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
752
■■■■■■■■
tiamo 2.2 (ProcessLab)
Drift value
Drift value for drift correction. This parameter is only shown for Drift cor-
rection = manual.
Range
0.0 ... 99.9 µL/min
Default value
0.0 µL/min
Stop volume
Determination stops when the specified volume has been dosed since the
start of the conditioning. The stop volume should be adjusted to the size of
the titration cell in order to prevent any overflow.
Titrando
Range
0.000000 ... 9999.99 mL
Default value
20,000 mL
Selection
off
Default value
off
Stop time
Determination stops when the specified time has elapsed since the start of
the conditioning.
Titrando
Range
1 ... 999999 s
Selection
off
Default value
off
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Range
0 ... 999 s
Default value
0 s
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Range
0 ... 99 s
Default value
0 s
Only start titration by a start command from a SEND command
on | off (Default value: off)
If this option is activated, then the titration will not be started until the Event
message Start titration is sent to the command with a SEND command.