2 General program functions
tiamo 2.2 (ProcessLab)
■■■■■■■■
33
Note
In the case of variables with Index {x}, the desired number 1…9 must be
entered for x (e.g. .EP{3}.ERC for the third endpoint).
If no index is specified, then the last index will be used automatically
(e.g. .EP.ERC for the last endpoint).
For the commands ANALOG IN, ANALOG OUT, DIGITAL IN, DIGI-
TAL OUT and STEPPING MOTOR, the syntax is 'Command-
name.Portname.Variabledesignation'.
Identifica-
tion
Description
Commands
.BLV
Blank value for the sensor used for
the command (only for ISE sensor)
and/or blank value calculated from
the calibration (for CAL LOOP Conc)
DET U, MET U,
SET U, STAT U,
MEAS U, MEAS
Conc, CAL LOOP
Conc, DOS U
.BP{x}.CNT
Intensity for the break point x (1…
9) in counts
MEAS Opt
.BP{x}.DME
Measured value differential for the
break point x (1…9)
MET
.BP{x}.ERC
ERC or first derivative for the break
point x (1…9)
DET
.BP{x}.MEA
Measured value for the break point
x (1...9) in the unit of the measured
value
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TEM
Temperature for the break point x
(1…9) in °C
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TIM
Time for the break point x (1…9) in
s
DET, MET, MEAS
(without T/Flow,
Ref, Spec)
.BP{x}.TRN
Transmission for the break point x
(1...9) in %
MEAS Opt
.BP{x}.VOL
Volume for the break point x (1…9)
in mL
DET, MET