5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
1009
5.6.4.17
STDADD auto
5.6.4.17.1
STDADD auto - Overview
Dialog window: Method
▶ STDADD auto ▶ Properties... ▶ STDADD auto - 'Command
name'
Command for Standard addition with automatic addition of the standard
addition solution from a dosing device in such a way that a constant poten-
tial difference results.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 857, 904, 905, 906, 907
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance:
Parameters
The parameters for the command STDADD auto are configured on the
following 3 tabs:
■
General/Hardware
Parameters for devices, dosing devices, sensors and stirrers.
■
Standard addition
Parameters for standard addition.
■
Measuring parameters
Parameters for setting the measurement procedure.
Standard addition curve
The standard addition curve with its associated evaluation data for a deter-
mination is shown in the program part Database in the dialog window
Calibration curve, which is opened with the menu item Determina-
tions
▶ Show calibration curve... or the symbol .
Command variables
The following command variables are generated in the method run by the
command STDADD auto and can be used in formulas under the designa-
tion 'Command name.Variable designation':