Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 835
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
819
Titrino
Selection
1 | 2 | diff.
Default value
1
Sensor
Selection of a sensor of type pH electrode from the list of sensors available
in the sensor table. The calibration data for the sensor will be adopted for
the determination.
Selection
Sensor name | pH electrode
Default value
pH electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.