Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 693
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
677
Sensor
Measuring input
Selection of the measuring input to which the sensor is connected.
Titrando
Selection
1 | 2
Default value
1
855
Selection
1
Default value
1
Titrino
Selection
1 | 2 | diff.
Default value
1
Sensor
Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode or ISE electrode from the list of sensors defined in the sensor table.
The calibration data for the sensor will be adopted for pH and ISE electrodes.
Selection
Sensor name | pH electrode | Metal electrode |
ISE electrode
Default value
Metal electrode
Temperature measurement
Type of temperature measurement.
Titrando, 855
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise the temperature entered manually under
Temperature on the Titration parameters tab will used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.