5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
832
■■■■■■■■
tiamo 2.2 (ProcessLab)
Note
If, when monitoring the dosing rate, the Lower limit is exceeded then
the actions Wait for [Continue] and Wait for limit ok are irrelevant
as the mean dosing rate becomes ever smaller during this time. The valid
range can never be reached again.
Monitoring Temperature
Monitoring Temperature
on | off (Default value: off)
If this option is switched on then the temperature will be monitored and any
exceeded limits will be entered in the measuring point list.
Lower limit
Lower limit of the temperature. When the temperature falls below this value,
the action Lower temperature limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
-20.0 °C
Titrino
Range
-170.0 ... 500.0 °C
Default value
-170.0 °C
Lower hysteresis
Lower hysteresis of the temperature. When the temperature once again
exceeds the lower limit by this hysteresis value, the action Temperature
lower limit ok is triggered.
Titrando, 855
Range
0.0 ... 150.0 °C
Default value
0.2 °C
Upper limit
Upper limit of the temperature. When the temperature exceeds this value,
the action Temperature upper limit exceeded is triggered.
Titrando, 855
Range
-20.0 ... 150.0 °C
Default value
150.0 °C