Page 1059: Cal meas conc
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
1043
5.6.5.6.2
CAL LOOP Conc - Properties
Dialog window: Method
▶ CAL LOOP Conc ▶ Properties... ▶ CAL LOOP Conc - 'Com-
mand name'
The number of standards and their concentration are defined in the dialog
window CAL LOOP Conc - 'Command name'.
Command name
Name of the command.
Input
25 characters
Standards
Number of standards
Selection of the number of standards with which the concentration calibra-
tion is to be carried out.
Selection
1 | 2 | 3 | 4 | 5
Default value
2
Unit conc.
Selection of the concentration unit for the standards.
Selection
mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L | µg/
mL | mg/mL
Default value
ppm
Conc. standard #
Enter the corresponding concentration for each standard used.
Range
0.001 ... 999999.999
Request for standard exchange
on | off (Default value: on)
If this option is enabled, a request for manual standard exchange is displayed
after each measurement. This option must be disabled for automatic solu-
tion exchange when calibrating with a Sample Processor.
5.6.5.7
CAL MEAS Conc
5.6.5.7.1
CAL MEAS Conc - Overview
Dialog window: Method
▶ CAL MEAS Conc ▶ Properties.. ▶ CAL MEAS Conc - 'Com-
mand name'
Measuring command for calibration with ISE electrodes, which is used
together with the command CAL LOOP Conc.