6 Configuration
tiamo 2.2 (ProcessLab)
■■■■■■■■
1303
Dosing rate Dosing port 2
Speed at which dosing is to be carried out at Dosing port 2. The maximum
dosing rate depends on the cylinder volume of the exchange unit used (see
below). When the function is carried out the speed will automatically be
reduced to the largest possible value.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Dosing rate Fill port
Speed at which dosing or filling is to be carried out at the Fill port. The
maximum dosing rate depends on the cylinder volume of the exchange unit
used (see below). When the function is carried out the speed will automat-
ically be reduced to the largest possible value.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
Dosing rate Special port
Speed at which dosing is to be carried out at the special port. The maximum
dosing rate depends on the cylinder volume of the exchange unit used (see
below). When the function is carried out the speed will automatically be
reduced to the largest possible value.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
The maximum dosing rate depends on the cylinder volume of the
dosing unit used (see below):
Cylinder volume
Maximum rate
2 mL
6.67 mL/min
5 mL
16.67 mL/min
10 mL
33.3 mL/min
20 mL
66.6 mL/min
50 mL
166.0 mL/min
Note
Enter lower rates for high-viscosity liquids.