5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
1007
Concentration
Concentration of addition solution.
Range
0.001 ... 999999.999
Default value
1.000
Device
Concentration unit of addition solution.
Selection
mol/L | mmol/L | µmol/L | ppm | g/L | mg/L |
µg/L | mg/mL | µg/mL | % | mEq/L
Default value
ppm
Dosing rate
Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.
Range
0.01 ... 166.00 mL/min
Selection
maximum
Default value
maximum
5.6.4.16.4
STDADD dos - Measuring parameters
Tab: Method
▶ STDADD ▶ Properties... ▶ Measuring parameters
Command name
Name of the command.
Input
25 characters
Measurement with drift control
The measurement is stopped as soon as the Signal drift or the Stop
measured value is reached or a defined Waiting time has elapsed.
Measured value drift
The measured value is only accepted if the signal drift defined here has been
fallen short of.
Range
0.1 ... 999.0 mV/min
Default value
10.0 mV/min
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.
Range
0 ... 999999 s
Default value