Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1090
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1074
■■■■■■■■
tiamo 2.2 (ProcessLab)
Identifica-
tion
Description
.EME
Final measured value (measured value after processing
of the command) in the unit of the measured value
.ENP
Electrode zero point of the sensor used for the command
(dimensionless)
.ETE
End temperature (temperature following processing of
the command) in °C
.EVT
End volume (total dosed volume at the end of the com-
mand) in mL
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point of
the measuring point list
.LP.DVT
dV/dt for the last measuring point in the measuring point
list (SET, KFT, STAT, DOS) or drift for the last measuring
point in in the measuring point list in µg/min (KFC)
.LP.EXx
External value × (1...3) for the last measuring point of the
measuring point list
.LP.MEA
Measured value for the last measuring point in the unit
of the measured value of the measuring point list
.LP.TEM
Temperature for the last measuring point of the meas-
uring point list in °C
.LP.TIM
Time in s until the last measuring point in the measuring
point list is reached
.LP.VOL
Volume for the last measuring point in the measuring
point list in mL
.MR.MRC
Correlation coefficient for mean dosing rate for the
whole range
.MR.MRS
Standard deviation for mean dosing rate for the whole
range in mL/min
.MR.MRT
Mean dosing rate for the whole range in mL/min.