4 analog out - analog output – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 1220
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1204
■■■■■■■■
tiamo 2.2 (ProcessLab)
If a command block originates from a different device, the Port descrip-
tion of which does not exist on the currently selected device, then the
affected lines will be presented in red script.
Functions
Moves the selected line up.
Moves the selected line down.
[New]
Opens the dialog window Analog output # in order to edit a new line
(see Chapter 5.6.11.4.4, page 1204).
[Properties]
Opens the dialog window Analog output # in order to edit the selected
line (see Chapter 5.6.11.4.4, page 1204).
[Delete]
Deletes the selected line.
5.6.11.4.4
ANALOG OUT - Analog output #
Dialog: Method
▶ ANALOG OUT ▶ Analog outputs ▶ Properties ▶ ANALOG OUT -
Analog output #
Port description
Selection of the analog output. The analog outputs of the selected device
are available. No port can be selected if not defined is selected in the
Device name field.
Type
Shows the type of the analog output.
Action
Selection
Set value | Set default
Default value
Set value
Set value
Writes the content of the field Value on the analog output.
Set default
Writes the content of the Default value field from the configuration on
the analog output.
Value
Mapped value which is to be written on the analog output. A fix value or a
formula (formula editor, CTRL F or right mouse button) can be entered.