Sensor
Selection of a sensor of type ISE electrode from the sensors configured in
the sensor table.
Selection
'Sensor name' | ISE electrode
Default value
ISE electrode
Temperature measurement
Type of temperature measurement.
Selection
continuous | automatic | off
Default value
automatic
continuous
A temperature sensor must be connected. The temperature will then be
measured continuously.
automatic
If a temperature sensor is connected then the temperature will be meas-
ured continuously. Otherwise, the temperature entered manually under
Temperature on the Measuring parameters tab will be used.
off
The temperature will not be measured. The temperature manually
entered under Temperature on the Measuring parameters tab will
be used.
Stirrer
Stirrer
Selection of the stirrer.
Selection
1 | 2 | 3 | 4 | off
Default value
1
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished.