5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1070
■■■■■■■■
tiamo 2.2 (ProcessLab)
Note
This formula applies only for using a single dosing device. The dosing rate
is equal to the flow rate in the case of tandem dosing.
If the required dosing rate exceeds the maximum dosing rate for the selected
exchange or dosing unit (or cylinder volume, respectively), then a larger cyl-
inder volume must be selected.
Specification of dosing rate and dosing time
You define the dosing rate with which dosing is to be carried out during a
particular time. As in the previous case, the filling times are not included in
the total, and dosing is prolonged accordingly.
Flow rate of a single dosing device
The cylinder volume and the filling time of the exchange or dosing unit must
be taken into account for the calculation of the effective flow rate.
Cylinder volume
Max. flow rate
Exchange unit
Dosing unit
1 mL
approx. 80 mL/h
---
2 mL
---
approx. 170 mL/h
5 mL
approx. 400 mL/h
approx. 430 mL/h
10 mL10 mL
approx. 800 mL/h
approx. 860 mL/h
20 mL
approx. 1.6 L/h
approx. 1.7 L/h
50 mL
approx. 4.0 L/h
approx. 4.3 L/h
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)