tiamo 2.2 (ProcessLab)
■■■■■■■■
793
5.6.3.6.10
KFC - Fix endpoint evaluation
Dialog window: Method
▶ KFC ▶ Properties... ▶ Additional evaluations ▶ Fix end-
point evaluation
▶ [New]/[Property] ▶ Fix endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | Time | Water
Default value
Measured value
Fix value
Value of the fix endpoint.
Measured value
Range
-2000.0 ... 2000.0 mV
Time
Range
0.0 ... 999999.9 s
Water
Range
0.0 ... 999999.9 µg
5.6.3.7
BRC
5.6.3.7.1
BRC - Overview
Dialog window: Method
▶ BRC ▶ Properties... ▶ BRC - 'Command name'
Command for coulometric titration for determining the bromine consump-
tion. Out of the amount of bromine consumed the bromine index (amount
of bromine consumed (in mg) per 100 g sample) and the bromine num-
ber (amount of bromine (in g) per 100 g sample) can be calculated.
Devices
This command can be executed with the following devices.
Titrandos: 851, 852
Appearance
The command has the following appearance:
without conditioning
with conditioning