Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 970
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
954
■■■■■■■■
tiamo 2.2 (ProcessLab)
Command variables
The following command variables are generated in the method run by the
command MEAS Conc and can be used in formulas under the designation
'Command name.Variable designation':
Identifica-
tion
Description
.BLV
Blank value of the sensor used for the command (only for
ISE sensors)
.BSY
Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY; invalid (variable not available) = Command has
not ever been started
.DBL
Total duration for the processing of the command in s
.EME
Final measured value (measured value after processing of
the command) in the unit of the measured value
.ENP
Electrode slope of the sensor used for the command (in
mV for ISE sensors)
.ETE
End temperature (temperature following processing of
the command) in °C
.FIN
Command status; 1 = Command has been ended at
least once, 0, invalid (variable not available) = Com-
mand has never been ended
.IME
Initial measured value (measured value before start con-
ditions are processed) in the unit of the measured value
.ITE
Initial temperature (temperature before start conditions
are processed) in °C
.LP.CAx
Calculated value × (1...3) for the last measuring point of
the measuring point list
.LP.EXx
External value × (1...3) for the last measuring point of the
measuring point list
.LP.MEA
Measured value for the last measuring point of the meas-
uring point list in the unit of the measured value
.LP.TEM
Temperature for the last measuring point of the measur-
ing point list in °C
.LP.TIM
Time in s until the last measuring point of the measuring
point list is reached
.MTE
Temperature measurement with sensor; 1 = on, 0 =
off)