2 General program functions
tiamo 2.2 (ProcessLab)
■■■■■■■■
37
Identifica-
tion
Description
Commands
.FIN
Command status; 1 = Command
has ended at least once, 0, inva-
lid (variable not available) = Com-
mand has never ended
All except START
and END
.FP{x}.CNT
Intensity for the fixed endpoint x
(1…9) in counts
MEAS Opt
.FP{x}.DME
Measured value differential for the
fixed endpoint x (1…9)
MET
.FP{x}.DVT
Drift for the fixed endpoint x (1…9)
in µg/min
KFC, BRC
.FP{x}.ERC
ERC for the fixed endpoint x (1…9) DET
.FP{x}.MEA
Measured value for the fixed end-
point x (1...9) in mV
DET, MET, SET,
KFT, STAT
.FP{x}.MEA
Measured value for the fixed end-
point x (1...9) in the unit of the
measured value
KFC, MEAS
(without T/Flow,
Conc, Ref, Spec)
.FP{x}.QTY
Measured value for the fixed end-
point x (1…9) in µg
KFC, BRC
.FP{x}.TEM
Temperature for the fixed endpoint
x (1…9) in °C
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Conc, Ref, Spec)
.FP{x}.TIM
Time in s until the fixed endpoint x
(1...9) is reached
DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without T/Flow,
Conc, Ref, Spec)
.FP{x}.TRN
Transmission for the fixed endpoint
x (1…9) in %
MEAS Opt
FP{x}.UGE
Voltage at the generator electrode
for the fixed endpoint # (1...9) in V.
In case of measurements with 756,
831: 0 = not defined, 1 = <14V,
2 = 14...28V, 3 = >28 V
KFC, BRC
.FP{x}.VOL
Volume for the fixed endpoint x
(1…9) in mL
DET, MET, SET,
KFT, STAT