5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
630
■■■■■■■■
tiamo 2.2 (ProcessLab)
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.
Stirrer
Stirrer
Selection of the stirrer.
Titrando, 855
Selection
1 | 2 | 3 | 4 | off
off
No stirrer will be used.
Stirring rate
Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.
Titrando, 855
Range
-15 ... 15
Default value
8
Switch off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched off automatically when
the command has finished. This parameter is displayed only for Titrandos
and 855.
Switch on/off automatically
on | off (Default value: on)
If this option is enabled, the stirrer will be switched on automatically at the
start of the command and will be switched off automatically at the end of
the command. This parameter is displayed only for 751, 785, 798 and 799.
5.6.3.3.5.3
MET Upol - Start conditions
Tab: Method
▶ MET Upol ▶ Properties... ▶ Start conditions
Command name
Name of the command.
Input
25 characters
The start conditions are processed in the listed sequence before the titration
is started.