8 brc - additional evaluations, Additional evaluations – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 821
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
805
Stabilizing time
Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.
Range
0 ... 999 s
Default value
0 s
Sample addition time
Waiting time between stop of conditioning and start of titration during
which the sample can be added.
Range
0 ... 99 s
Default value
0 s
Only start titration by a start command from a SEND command
on | off (Default value: off)
If this option is activated, then the titration will not be started until the Event
message Start titration is sent to the command with a SEND command.
Note
Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titration
will be ignored.
If this command is already started (e.g. via Automatic start after sam-
ple addition), the SEND command will not have any effect.
Show measured value during conditioning
on | off (Default value: off)
If this option is enabled, then the measured value will be shown in the sub-
window Live display of the program part Workplace, in addition to the
drift and the time.
5.6.3.7.8
BRC - Additional evaluations
Tab: Method
▶ BRC ▶ Properties... ▶ Additional evaluations
Command name
Name of the command.
Input
25 characters
The following additional method for evaluation of the titration curves can
be activated and defined on this tab: