Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 957
■■■■■■■■■■■■■■■■■■■■■■
5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
941
Min. waiting time
The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift continues
to be checked while the waiting time is elapsing.
Titrando, 855, 867
Range
0 ... 999999 s
Default value
0 s
Max. waiting time
If the signal drift has not yet been reached, then the measured value will be
accepted when the maximum waiting time has elapsed. If the waiting time
has not been newly entered then a waiting time that is suitable for the drift
will be calculated automatically according to the following equation:
Titrando, 855, 867
Range
0 ... 999999 s
Default value
215 s
Titrino
Range
0 ... 9999 s
Default value
215 s
Measuring interval
Time interval for entering a measuring point in the measuring point list.
Titrando, 855, 867
Range
0.1 ... 999999.0 s (Increment: 0.1)
Default value
2.0 s
Titrino
Range
0.08 ... 16200 s (Increment: 0.08)
Default value
2.0 s
Stop meas. value
Stops when the preset measured value has been reached since the start of
the measurement.
Titrando, 855, 867
Range
-20.0 ... 150.0 °C
Selection
off
Default value
off