6 meas ipol - fix endpoint evaluation, 6 meas upol, 1 meas upol - overview – Metrohm tiamo 2.2 (ProcessLab) User Manual
Page 938: Meas upol
5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
922
■■■■■■■■
tiamo 2.2 (ProcessLab)
5.6.4.5.6
MEAS Ipol - Fix endpoint evaluation #
Dialog window: Method
▶ MEAS Ipol ▶ Properties... ▶ Additional evaluations ▶ Fix
endpoint evaluation #
Quantity
Selection of the fixed measured quantity to which the associated values for
the other quantities are to be interpolated from the list of measured points.
Selection
Measured value | time
Default value
Measured value
Fix value
Value of the fixed endpoint.
Measured value (Titrando, 867, 855)
Range
-2000.0 ... 2000.0 mV
Measured value (Titrino)
Range
-2000 ... 2000 mV
Time
Range
0.0 ... 999999.9 s
5.6.4.6
MEAS Upol
5.6.4.6.1
MEAS Upol - Overview
Dialog window: Method
▶ MEAS Upol ▶ Properties... ▶ MEAS Upol - 'Command
name'
Command for amperometric measurements with selectable polarization
voltage.
Devices
This command can be executed with the following devices:
Titrando: 808, 809, 835, 836, 841, 852, 857, 888, 890, 901, 904, 905,
906, 907
Titrino: 702, 716, 718, 719, 720, 721, 736, 751, 758, 785, 794, 798, 799
pH/Ion - Meter: 867
Robotic Titrosampler: 855
Appearance
The command has the following appearance: