5 Method
tiamo 2.2 (ProcessLab)
■■■■■■■■
855
Titrando, 855
Range
0.0 ... 2000.0 mV
Default value
0.2 mV
Action
Selection of the event to be triggered if the upper or lower limit is exceeded:
Selection
Stop determination | Stop command (Titrando,
855) | Wait for [Continue] | Wait for limit ok |
none
Default value
none
Stop determination
The running STAT command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (Titrando, 855)
The running STAT command is quit, then the next command is executed.
Wait for [Continue]
Reagent dosing in the current STAT command is interrupted and a mes-
sage appears. As soon as the monitored measured value is again within
the limits (including hysteresis) reagent dosing can be resumed by press-
ing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.
Monitoring Dosing rate
Monitoring Dosing rate
on | off (Default value: off)
If this check box is activated, then the averaged dosing rate will be monitored
and any exceeded limits will be entered in the measuring point list. This
monitoring is not started until 10 s after the start of the titration.
Lower limit
Lower limit of the dosing rate. When the dosing rate falls below this value,
the action Dosing rate lower limit exceeded will be triggered.
Titrando, 855
Range
0.00 ... 166.00 mL/min
Default value
0.00 mL/min