beautypg.com

Spectra Precision FAST Survey Reference Manual User Manual

Page 330

background image

p328

Referring to the graphic below, which illustrates a stockpile, the “Final” point range could be defined as ALL points,

and the “Original” point range could be defined as the base points of the stockpile or points 1-15.

So using the settings above in the Volume routine, you would verify your two point ranges, then click OK. For

example, the “Original” surface point range would be set as follows using the “Define” button for “Original” prior to

clicking OK.

Clicking OK moves forward to a shrink and swell factor screen. Only if you are set to English units do you get the

option for tons based on a density factor. In metric units, the volume is presented in cubic meters. You can create a new

surface TIN file that merges the original surface with the final surface within the selected or assumed perimeter. If you

do not have an inclusion perimeter, then the largest "convex" figure defined by all final surface points and entities

becomes the default perimeter.

Clicking OK continues on to the Volume Report.