MTS Series 793 User Manual
Page 329

Warning:
Changes in Function Generator values made with hydraulic pressure on can result in
unexpected actuator movement.
A moving actuator can injure anyone in its path.
Always clear the actuator area before changing values.
2. In the Function Generator panel:
a) Select the Channel to be programmed.
b) Select a Control Mode of displacement.
c) Select a Command Type of Cyclic.
d) Set a Target Setpoint of midstroke.
Typically, the actuator should cycle around midpoint.
e) Set an Amplitude (±) that will move the actuator through about 70% of its full displacement.
f) Set a Frequency of 0.25–0.5 Hz.
Displayed ranges for Target Setpoint, Amplitude, and Frequency are saved and are restored when
returning to a previously programmed channel.
g) Select a Wave Shape of Sine.
h) As needed, select a Compensator.
Warning:
Clicking the Run button will put actuators in motion.
A moving actuator can injure anyone in its path.
Always clear the test area before pressing the run button.
3. In the Station Manager window’s Station Controls panel:
a) Click Reset to clear interlocks.
b) Apply station power.
c)
Click
to start the program.
MTS Series 793 Control Software 329
Using Station Manager to Run Tests