Program Hold Interlock
TEDS Sensors
About TEDS Sensors
Note:
Only controllers that use Series 494 Hardware support the use of MTS TEDS (transducer electronic
data sheet) sensors.
TEDS sensors have built-in memory chips that store basic TEDS information (manufacturer, model, number,
and serial number).
When you connect an initialized TEDS device to a conditioner, the Station Manager application reads the
basic TEDS information from the device, and a sensor assignment window appears.
The sensor file list in this window only lists sensor files that match the basic TEDS information, conditioner
type, and dimension. The sensor assignment window contains the same sensor assignment list found in the
Station Setup window.
TEDS devices
Available TEDS devices include:
•
MTS sensors that include basic TEDS information saved in an integral TEDS chip that is built into the
sensor.
•
MTS TEDS Transducer ID modules that include a chip that stores basic TEDS information for a specific
sensor. The TEDS module connects to the sensor.
About Calibration Data for TEDS Sensors
There are two methods that you can use to assign calibration data to a TEDS Sensor:
Station Setup
If the sensor is already connected to the system, you can use the Station Setup window to assign calibration
data to the TEDS sensor.
218 MTS Series 793 Control Software
Station Setup