Description
Callout
Select a sensor input
1
Select a sensor file
2
Assign the file
3
3. For bridge-type sensors (for example, load cells), perform a shunt calibration check before running a test.
Causes
The Station Manager application changes the sensor-file assignment to
•
The dimension or conditioner type settings in the sensor file do not match the settings in the Station Setup
window. This can occur if you modify the sensor file or change the station configuration.
•
The Station Manager application cannot locate the assigned sensor file. This can occur if the sensor file
was renamed, deleted, or moved to another directory.
Actions
An invalid sensor-file assignment results in the following actions:
•
When the station loads, the message log lists any invalid sensor-file assignments.
•
The Sensor File selection in the Station Setup window is appended with
•
If the sensor-file assignment is set to
•
If the active mode sensor input is in an
Sensor File Save Options
The Station Setup window provides the following sensor file save options:
Description
Item
Click Save to save any changes in the Sensor, Calibration, and Shunt tabs in
the currently assigned sensor file.
Save