beautypg.com

Specimen installation, About manual command – MTS Series 793 User Manual

Page 319

background image

A C-Stop interlock results in the following actions:

A Program Stop interlock occurs.

The Station Manager C-Stop indicator turns on.

The Station Manager application logs a message.

The control channel executes the defined C-Stop action.

The C-Stop parameters are defined on the Channel Options window > C-Stop tab

Note:

C-Stop differs from Ramp To and Hold At Level actions because it does not require the Function
Generator to be running.

Specimen Installation

About Manual Command

This window provides manual control of channels. Use this control to position the actuator during specimen
installation.

Important:

To help protect specimens and fixturing, you may want to use the Station Builder application to create
a Channel Limited Channel (CLC) control mode that you can use as the Manual Command control
mode. The CLC control mode is typically set up as displacement control with a load signal input with
limit settings that can limit the amount of force that can be applied during specimen installation.

Access

Station Controls panel toolbar >

MTS Series 793 Control Software 319

Using Station Manager to Run Tests